+ <h2><b>profile/colverify</b></h2>
+ <h3>Summary</h3>
+ Verify a color transform by comparing CIE measurement values from
+ two test charts. The charts can be any suitably formatted CGATS or <a
+ href="File_Formats.html#.ti3">.ti3</a> format that contains
+ corresponding XYZ, Lab or spectral values. The overall average and
+ worst case delta E will be reported, as well as the worst 10% and
+ best 90% of values.<br>
+ <h3>Usage Summary</h3>
+ <tt><small>colverify&nbsp; [-options] target.ti3 measured.ti3<br>
+ &nbsp;-v
+ [n]&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+ Verbose mode, n &gt;= 2 print each value<br>
+ &nbsp;-n&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+Normalise
+
+
+ each files reading to white Y<br>
+ &nbsp;-N&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+Normalise
+
+
+ each files reading to white XYZ<br>
+ &nbsp;-D&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+Use
+
+
+ D50 100.0 as L*a*b* white reference<br>
+ &nbsp;-c&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+&nbsp;
+
+
+ &nbsp; Show CIE94 delta E values<br>
+ &nbsp;-k &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+ &nbsp; &nbsp; Show CIEDE2000 delta E values<br>
+ &nbsp;-s &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+ &nbsp; &nbsp; Sort patch value by error<br>
+ &nbsp;-w&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+&nbsp;
+
+
+ &nbsp; create VRML vector visualisation (measured.wrl)<br>
+ </small></tt><tt><small>&nbsp;-W
+ &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; &nbsp;
+ &nbsp; create VRML marker &amp; vector visualisation
+ (measured.wrl)</small></tt><tt><br>
+ </tt><tt> </tt><tt><small>&nbsp;-x&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+&nbsp;
+
+
+ &nbsp; Use VRML axes<br>
+ &nbsp;</small></tt><tt><small><small>-f
+ [illum]&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Use Fluorescent
+ Whitening Agent compensation [opt. simulated inst. illum.:<br>
+ &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+
+
+ M0, M1, M2, A, C, D50 (def.), D50M2, D65, F5, F8, F10 or
+ file.sp]<br>
+ &nbsp;-i illum&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+ Choose illuminant for computation of CIE XYZ from spectral
+ data &amp; FWA:<br>
+ &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+
+
+ A, C, D50 (def.), D50M2, D65, F5, F8, F10 or file.sp<br>
+ &nbsp;-o observ&nbsp;&nbsp;&nbsp; &nbsp;&nbsp;&nbsp; Choose
+ CIE Observer for spectral data:<br>
+ </small><small>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;
+
+
+ 1931_2 </small><small>(def.)</small><small>, 1964_10, S&amp;B
+ 1955_2, shaw, J&amp;V 1978_2<br>
+ &nbsp;-L profile.icm&nbsp;&nbsp; Skip any first file out of
+ profile gamut patches<br>
+ &nbsp;-X file.ccmx&nbsp;&nbsp;&nbsp;&nbsp; Apply Colorimeter
+ Correction Matrix to second file<br>
+ </small> &nbsp;<i>target.ti3</i>&nbsp;&nbsp;&nbsp;&nbsp; &nbsp;
+ Target or reference patch data file<br>
+ &nbsp;<i>measured.ti3</i> &nbsp; &nbsp; Measured or actual patch
+ data file</small></tt><br>
+ <h3>Usage Details and Discussion</h3>
+ <b> colverify</b> provides a way of verifying how well a color
+ transformation (such a proofing) performs.<br>
+ <br>
+ The <b>-v</b> flag prints out extra information during the
+ checking, and prints each patch value, rather than just a summary.<br>
+ <br>
+ The <b>-n</b> flag causes the two sets of values to be normalized
+ to the Y value of white for each set before comparison. White is
+ assumed to be the patch with the largest Y value.<br>
+ <br>
+ The <b>-N</b> flag causes the two sets of values to be normalized
+ to the XYZ of white for each set before comparison. White is assumed
+ to be the patch with the largest Y value.<br>
+ <br>
+ The <b>-D</b> flag causes the white reference point for the
+ conversion to L*a*b* to be D50 with a Y value of 100%. By default
+ the sample with the largest Y value is found, and the L*a*b* white
+ reference scaled to have that Y value. This allows sensible delta E
+ values when comparing absolute color values, such as those from
+ emission or display measurements.<br>
+ <br>
+ The <b>-c</b> option causes the differences between the test values
+ and the profile prediction of the color for each device value to be
+ displayed in CIE94 delta E, rather than plain L*a*b* delta E. CIE94
+ delta E has a closer correspondence with perceived color differences
+ than the default CIE76 delta E values.<br>
+ <br>
+ The <b>-k</b> option causes the differences between the test values
+ and the profile prediction of the color for each device value to be
+ displayed in CIEDE2000 delta E, rather than plain L*a*b* delta E.
+ CIEDE2000 delta E has a closer correspondence with perceived color
+ differences than either CIE76 or CIE94 delta E values.<br>
+ <br>
+ If the <b>-s </b>flag is used in combination with the <b>-v</b>
+ flag, then the test point by test point output will be sorted from
+ worst to best.<br>
+ <br>
+ The <b>-w</b> creates a <a href="File_Formats.html#VRML">VRML</a>
+ 3D visualization of the differences between the test points in D50
+ L*a*b* space, each difference being shown as a line vector. If the <span
+ style="font-weight: bold;">-W</span> flag is used, then the target
+ and measured values will also be marked by a small sphere. This can
+ be used to visualize the placement of values in a .ti3 (or other
+ CGATS file) by using the same file for both "target" and "measured"
+ values.<br>
+ <br>
+ The <b>-x</b> flag adds Lab axes to the VRML output.<br>
+ <br>
+ The <b>-f</b> flag enables Fluorescent Whitening Agent (FWA)
+ compensation. This only works if spectral data is available and, the
+ instrument is not UV filtered.&nbsp; FWA compensation adjusts the
+ spectral samples so that they appear to have been measured using an
+ illuminant that has a different level of Ultra Violet to the one the
+ instrument actually used in the measurement. The optional
+ illumination parameter allows specifying a standard or custom
+ illumination spectrum to be used as the similated instrument
+ illuminant, overriding the default <b>D50</b> or CIE computation
+ illuminant used for FWA (see <b>-i</b> below<b>). </b>See <a
+ href="colprof.html#f">colprof -f</a> for a fuller explanation. The
+ same value should be used as was used during the creation of the
+ profile.<br>
+ <br>
+ The <b>-i</b> flag allows specifying a standard or custom
+ illumination spectrum, applied to the spectral test point values to
+ compute CIE tristimulus values. <b>A</b>, <b>D50</b>, <b>D50M2,
+ D65</b>, <b>F5</b>, <b>F8</b>, <b>F10</b> are a selection of
+ standard illuminant spectrums, with <b>D50</b> being the default.
+ If a filename is specified instead, it will be assumed to be an
+ Argyll specific <a href="File_Formats.html#.sp">.sp</a> spectrum
+ file. If FWA compensation is used during measurement, this
+ illuminant will be used by default as the simulated instrument
+ illuminant. The same value should be used as was used during the
+ creation of the profile.<br>
+ <br>
+ The <b>-o</b> flag allows specifying a tristimulus observer, and is
+ used to compute PCS (Profile Connection Space) tristimulus values.
+ The following choices are available:<br>
+ <b>&nbsp; 1931_2</b> selects the standard CIE 1931 2 degree
+ observer. The default.<br>
+ &nbsp; <b>1964_10</b> selects the standard CIE 1964 10 degree
+ observer.<br>
+ &nbsp; <b>1955_2</b> selects the Stiles and Birch 1955 2 degree
+ observer<br>
+ &nbsp; <b>1978_2 </b>selects the Judd and Voss 1978 2 degree
+ observer<br>
+ &nbsp; <b>shaw</b> selects the Shaw and Fairchild 1997 2 degree
+ observer<br>
+ <br>
+ The same parameter value should be used as was used during the
+ creation of the profile.<br>
+ <br>
+ The <b>-L</b><b> profile.icm</b> parameter causes colverify to ignore
+ any patches from the first file that are out of gamut according to
+ the profile. This can be useful in evaluating accuracy without
+ tainting it by impossible to reach colors.<br>
+ <br>
+ The <b>-X</b><b> file.ccmx</b> option applies the given 3x3
+ calibration matrix to the values from the second file before doing
+ any verification. This can be useful in evaluating the effects of a
+ calibration matrix on raw colorimeter values, against reference
+ values measured using a spectrometer.<br>
+ <br>
+ If both CIE and spectral values are present in the input files, the
+ CIE values will be used by default. Using the <span
+ style="font-weight: bold;">-i</span>, <span style="font-weight:
+ bold;">-o</span> or <span style="font-weight: bold;">-f</span>
+ flag will force spectral values to be used. The the <span
+ style="font-weight: bold;">-i</span>, <span style="font-weight:
+ bold;">-o</span> or <span style="font-weight: bold;">-f</span>
+ flags will apply to both the target and measured input files.<br>
+ <br>
+ <br>
+ <br>
+ <br>
+ <br>
+ <br>
+ </body>