spectro/spec2cie
Summary
Convert reflective spectral .ti3
readings into CIE XYZ and D50 L*a*b* readings. Apply FWA, plot
spectrums.
Usage
spec2cie [options] input.ti3
output.ti3
-v
Verbose mode
-I illum
Override actual instrument illuminant in .ti3
file:
A, C, D50, D50M2, D65, F5, F8, F10 or
file.sp
(only used in conjunction with -f)
-f [illum] Use
Fluorescent Whitening Agent compensation [simulated inst.
illum.:
M0, M1, M2, A, C, D50 (def.), D50M2, D65, F5, F8, F10 or
file.sp]
-i illum
Choose illuminant for computation of CIE XYZ
from spectral data & FWA:
A, C, D50 (def.), D50M2, D65, F5,
F8, F10 or file.sp
-o observ
Choose CIE Observer for spectral data:
1931_2 (def.),
1964_10, S&B 1955_2, shaw, J&V 1978_2
-n
Don't
output
spectral values
-p
Plot each values spectrum
input.ti3
Measurement
file
output.ti3
Converted measurement file
Comments
This program takes the (usually reflective) spectral data in a .ti3
file, converts them to XYZ and D50 L*a*b* and fills the XYZ_[XYZ]
and LAB_[LAB] columns in the output .ti3 file. If the columns
XYZ_[XYZ] and/or LAB_[LAB] are missing in the input file, they are
added to the output file.
If FWA correction is applied, then the spectra saved will be FWA
corrected, as well as the XYZ and D50 L*a*b* values.
All other columns are copied from the input to the output .ti3 file.
Setting an actual instrument/simulated instrument/CIE illuminant
only has an effect for reflective/transmissive spectra. Emissive
spectra don't have an illuminant, and will simply be converted to
XYZ using the given observer.
The -I parameter allows specifying a an
illumination spectrum for the actual instrument illuminant,
overriding that computed from the type of instrument (recorded in
the .ti3 file). This is rarely used. This parameter is can only be
used in combination with the -f
flag. If a filename is specified instead, it will be assumed to be
an Argyll specific .sp custom
spectrum file. Illuminant details are:
A CIE
tungsten filament lamp 2848K
D50 CIE daylight 5000K
D65 CIE daylight 6500K
F5 CIE Fluorescent
6350K, CRI 72
F8 CIE Fluorescent
5000K, CRI 95
F10 CIE Fluorescent
5000K, CRI 81
The -f flag enables Fluorescent Whitening
Agent (FWA) compensation. This only works if spectral data is
available and, the instrument is not UV filtered. FWA
compensation adjusts the spectral samples so that they appear to
have been measured using an illuminant that has a different level of
Ultra Violet to the one the instrument actually used in the
measurement. The optional illumination parameter allows specifying a
standard or custom illumination spectrum to be used as the similated
instrument illuminant, overriding the default D50 or CIE
computation illuminant used for FWA (see -i below). See
colprof -f for a fuller explanation.
The -i parameter allows specifying a
standard or custom illumination spectrum, applied to spectral .ti3
data to compute PCS (Profile Connection Space) tristimulus values. A,
D50, D65, F5, F8, F10 are a
selection of standard illuminant spectrums, with D50 being
the default. If a filename is specified instead, it will be assumed
to be an Argyll specific .sp
custom spectrum file. This only works if spectral data is available.
Illuminant details are:
A CIE
tungsten filament lamp 2848K
D50 CIE daylight 5000K
D65 CIE daylight 6500K
F5 CIE Fluorescent
6350K, CRI 72
F8 CIE Fluorescent
5000K, CRI 95
F10 CIE Fluorescent
5000K, CRI 81
Custom illuminants are most often used when a viewing booth or
other known viewing conditions is going to be used to view results.
Other illuminant reference files could be created using a suitable
measuring instrument such as a spectrolino, or an eyeone using spotread, although such instruments do
not themselves provide the necessary response down to Ultra Violet
that is needed for accurate operation of Fluorescent Whitening Agent
compensation. The best way of measuring a custom illuminant is to
use illumread, since it uses a special
method to estimate the illuminant UV in a way that complements FWA
compensation. (See the discussion above for the -f flag).
Note that if an illuminant other than D50 is chosen, the resulting
ICC profile will not be standard, and may not work perfectly with
other profiles that that use the standard ICC D50 illuminant,
particularly if the absolute rendering intent is used. Profiles
should generally be linked with other profiles that have the same
illuminant and observer.
The -o flag allows specifying a tristimulus
observer, and is used to compute PCS (Profile Connection Space)
tristimulus values. The following choices are available:
1931_2 selects the standard CIE 1931 2 degree observer
(the default).
1964_10 selects the standard CIE 1964 10 degree
observer.
1955_2 selects the Stiles and Birch 1955 2 degree
observer
1978_2 selects the Judd and Voss 1978 2 degree
observer
shaw selects the Shaw and Fairchild 1997 2 degree
observer
Note that if an observer other than 1931 2 degree is chosen, the
resulting ICC profile will not be standard, and cannot be freely
interchanged with other profiles that that us the standard 1931 2
degree observer. Profiles should only be linked with other profiles
that have the same illuminant and observer.
The -n flag
disables the output of the spectral values. If just the XYZ and D50
L*a*b* values are needed, this makes for a smaller, easier to read
file.
The -p flag
causes each spectrum to be plotted. If FWA compensation is used,
then the before/after compensation spectruma will be shown.