summaryrefslogtreecommitdiff
path: root/doc/i1proDriver.html
blob: 0894f59d28c6c413928925187bf8d818d6482943 (plain)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
<!DOCTYPE html PUBLIC "-//W3C//DTD HTML 4.01 Transitional//EN">
<html>
<head>
  <title>The i1pro Driver</title>
  <meta http-equiv="content-type"
 content="text/html; charset=ISO-8859-1">
  <meta content="Graeme Gill" name="author">
</head>
<body>
<h2 style="text-decoration: underline; font-weight: bold;">How can I
have confidence in the i1pro Driver ?<br>
</h2>
A question that has been asked is : "<span style="font-weight: bold;">You've
written your own driver for the Eye-One Pro. How can I have confidence
that the measurements are accurate, and will match those made with the
original manufacturers driver </span>?"<br>
<br>
This is a quite reasonable question. The following attempts to answer
it.<br>
<h4 style="text-decoration: underline;">Why does Argyll use it's own
i1pro driver ?</h4>
Primarily because the Original Manufacturers Driver (OMD) isn't
available for all the platforms that ArgyllCMS supports (Linux in
particular). A side benefit is that it's possible to tweak many of the
driver parameters for slightly better results and more flexibility. It
has also helped in understanding the characteristics and limitations of
such instruments.<br>
<h4 style="text-decoration: underline;">Does it match the OMD ?</h4>
In principle the behaviour should be very similar. While the Argyll
driver has been written from scratch, it does use exactly the same
calibration values from<br>
inside the instrument, and attempts to use the calibration values and
process the raw instrument readings in an equivalent manner to that of
the OMD.<br>
<br>
But the proof of the pudding is in the measuring, so to actually verify
this, the following experiment was conducted:<br>
<br>
The Argyll version used was V1.2.0<br>
<br>
The Macbeth 24 patch ColorChecker was used as a sample target. For each
patch (and the calibration tile), the following steps were performed:<br>
<br>
1) Place the instrument on the calibration tile.<br>
<br>
2) Use Argyll spotread to calibrate the Argyll driver.<br>
<br>
3) Change drivers to the OMD.<br>
<br>
4) Use the OMD to calibrate the instrument.<br>
<br>
5) Move the instrument to the patch on the ColorChecker.<br>
<br>
6) Read the color using the OMD.<br>
<br>
7) Change the back to the Argyll driver.<br>
<br>
8) Using the calibration made in step 2), read the color using Argyll.<br>
<br>
Each calibration or reading was performed 15 seconds from the previous
one, to put the instrument lamp in a repeatable state.<br>
The instrument was kept in exactly the same position for calibration
and patch measurement with the two drivers.<br>
(The whole idea is to reduce all other sources of error, other than the
driver itself.)<br>
<br>
This measurement was repeated just once for each patch + the
calibration tile. This was done in one run, and the readings were not
specially selected.<br>
<h4 style="text-decoration: underline;">Results:</h4>
The following D50 L*a*b* values were recorded for each measurement:<br>
<br>
A) &nbsp;&nbsp; The OMD internally calculated L*a*b* value<br>
B) &nbsp;&nbsp; The L*a*b* value calculated by Argyll from the OMD
spectral values.<br>
C) &nbsp;&nbsp; The L*a*b* value calculated from the Argyll measured
spectral values.<br>
D) &nbsp;&nbsp; The L*a*b* value calculated from the Argyll
Hi-Resolution mode measured spectral values.<br>
<br>
<span style="text-decoration: underline;">A is compare to B, to check
that the spectral to standard observer calculations are equivalent.</span><br>
<br>
&nbsp;&nbsp;&nbsp; The result was an average Delta E (CIE76) of 0.006,
with a maximum of 0.012.<br>
<br>
&nbsp;&nbsp;&nbsp; This shows that there is very close agreement in the
way spectral values are converted to XYZ and L*a*b*.<br>
<br>
<span style="text-decoration: underline;">B is compared to C to check
that the Argyll driver behaves the same as the OMD.</span><br>
<br>
&nbsp;&nbsp;&nbsp; The result was an average Delta E (CIE76) of 0.028,
with a maximum of 0.051.<br>
<br>
&nbsp;&nbsp;&nbsp; This shows that the OMD and Argyll driver are in
close agreement in spectral measurement.<br>
&nbsp;&nbsp;&nbsp; This error is an order of magnitude smaller than
uniformity induced errors typical in the media being measured.<br>
<br>
<span style="text-decoration: underline;">A is compared to C to check
that the Argyll driver and spectral to XYZ differences don't compound.</span><br>
<br>
&nbsp;&nbsp;&nbsp; The result was an <span style="font-weight: bold;">average</span>
Delta E (CIE76) of <span style="font-weight: bold;">0.026</span>, with
a <span style="font-weight: bold;">maximum</span> of <span
 style="font-weight: bold;">0.048</span>.<br>
<br>
&nbsp;&nbsp;&nbsp; Rather than compounding, any spectral to XYZ
differences tend to cancel
out slightly. This is the <span style="font-weight: bold;">bottom line</span>
experimental difference between
the two drivers. The actual underlying difference may in fact be less
than this, but it would be necessary to do multiple test runs to
filter out experimental error.<br>
<br>
<span style="text-decoration: underline;">C is compare to D to check
that the Argyll Hi-Resolution mode is behaving reasonably.</span><br>
<br>
&nbsp;&nbsp;&nbsp; The result was an average Delta E (CIE76) of 0.158,
with a maximum of 0.353.<br>
<br>
&nbsp;&nbsp;&nbsp; Because the ColorChecker samples have relatively
smooth reflectance spectra, it can be expected that<br>
&nbsp;&nbsp;&nbsp; the normal and Hi-Res mode results should be fairly
similar. And indeed, this is the case. The biggest<br>
&nbsp;&nbsp;&nbsp; differences are for patches
with the largest spectral transitions in them, which is to be expected
as the<br>
&nbsp;&nbsp;&nbsp; Hi-Res measurement more
closely follows the spectral shape, while the differences for
spectrally flat<br>
&nbsp;&nbsp;&nbsp; patches is neglegable, since both can follow the
spectral shape well.<br>
<br>
Example Yellow-Green Patch, Hi-Res &amp; Normal spectrum:<br>
<img style="width: 709px; height: 259px;"
 alt="Yellow-Green patch, Hi-Res vs. Normal" src="YellowGreen.jpg"><br>
<br>
<h4 style="text-decoration: underline;">Conclusions:</h4>
The experimental average difference of <span style="font-weight: bold;">0.026</span>
Delta E76 shown above provides evidence that despite using a completely
different instrument driver to that supplied with the instrument, the
ArgyllCMS Eye-One pro measurement values have comparable accuracy, and
can be relied upon to match measurements made using the original
manufactures driver.<br>